{"schemaVersion":1,"url":"https://cve.openhack.com/cve/CVE-2025-64298","id":"CVE-2025-64298","title":"Mirion Medical EC2 Software NMIS BioDose Incorrect Permission Assignment for Critical Resource","description":"NMIS/BioDose V22.02 and previous version installations where the embedded Microsoft SQLServer Express is used are exposed in the Windows share accessed by clients in networked installs. By default, this directory has insecure directory paths that allow access to the SQL Server database and configuration files, which can contain sensitive data.","descriptionLong":"NMIS/BioDose V22.02 and previous version installations where the embedded Microsoft SQLServer Express is used are exposed in the Windows share accessed by clients in networked installs. By default, this directory has insecure directory paths that allow access to the SQL Server database and configuration files, which can contain sensitive data.","published":"2025-12-02T21:05:38.266Z","updated":"2026-09-25T23:10:00.463Z","reserved":"2025-11-11T20:56:52.837Z","status":"Analyzed","rejected":false,"assigner":"icscert","metric":{"version":"4.0","score":8.6,"vector":"CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N","rating":"high","source":"icscert"},"metrics":[{"version":"4.0","score":8.6,"vector":"CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N","rating":"high","source":"icscert"},{"version":"4.0","score":8.6,"vector":"CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","rating":"high","source":"hq.dhs.gov"},{"version":"3.1","score":8.4,"vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","rating":"high","source":"icscert"},{"version":"3.1","score":8.4,"vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","rating":"high","source":"hq.dhs.gov"},{"version":"3.1","score":7.5,"vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N","rating":"high","source":"NVD"}],"rating":"high","products":["Mirion Medical EC2 Software NMIS BioDose"],"affectedProducts":[{"vendor":"Mirion Medical","product":"EC2 Software NMIS BioDose","defaultStatus":"unaffected","versions":[{"lessThan":"23.0","status":"affected","version":"0","versionType":"custom"},{"status":"unaffected","version":"23.0"}],"platforms":[]}],"packages":[],"cwes":["CWE-732"],"aliases":[],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-336-01","tags":["Third Party Advisory"]}],"patchReferenced":false,"fixAvailable":false,"exploitReferenced":false,"kev":false,"kevAdded":null,"kevDue":null,"kevEntry":null,"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"total"},"epss":0.00251,"epssPercentile":0.16711,"sources":[{"name":"CVE List","url":"https://www.cve.org/CVERecord?id=CVE-2025-64298","ok":true},{"name":"NVD","url":"https://nvd.nist.gov/vuln/detail/CVE-2025-64298","ok":true,"note":"Analyzed"},{"name":"OSV","url":"https://osv.dev/vulnerability/CVE-2025-64298","ok":false,"note":"no open-source package data"},{"name":"CISA KEV","url":"https://www.cisa.gov/known-exploited-vulnerabilities-catalog","ok":true,"note":"not listed"},{"name":"EPSS","url":"https://api.first.org/data/v1/epss?cve=CVE-2025-64298","ok":true,"note":"scored 2026-11-19"}],"aiTitle":false,"tech":[],"critical":[],"attention":null}