{"schemaVersion":1,"url":"https://cve.openhack.com/cve/CVE-2025-14097","id":"CVE-2025-14097","title":"Remote Code Execution Vulnerability in Radiometer Products","description":"A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means.  The issue is caused by a weakness in the analyzer’s application software.                    ","descriptionLong":"A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means.  The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\nRequired Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution:\nCustomers should ensure the following:\n•  The network is secure, and access follows best practices.\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                     \nExploit Status:\n\n\nResearchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.","published":"2025-12-17T12:36:03.779Z","updated":"2025-12-17T14:41:13.252Z","reserved":"2025-12-05T10:50:03.683Z","status":"PUBLISHED","rejected":false,"assigner":"Radiometer","metric":{"version":"3.1","score":7.2,"vector":"CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H","rating":"high","source":"Radiometer"},"metrics":[{"version":"3.1","score":7.2,"vector":"CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H","rating":"high","source":"Radiometer"}],"rating":"high","products":["Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers","Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers","Radiometer Medical Aps AQT90 FLEX Analyzers"],"affectedProducts":[{"vendor":"Radiometer Medical Aps","product":"ABL90 FLEX and ABL90 FLEX PLUS Analyzers","defaultStatus":"unaffected","versions":[{"status":"affected","version":"Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems","versionType":"ABL90 DMS (Data Management system)"},{"status":"unaffected","version":"Application software versions >= 3.5MR11 with Windows 10 operating system","versionType":"ABL90 DMS (Data Management system)"}],"platforms":[]},{"vendor":"Radiometer Medical Aps","product":"ABL800 BASIC and ABL800 FLEX Analyzers","defaultStatus":"unaffected","versions":[{"status":"affected","version":"Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems","versionType":"ABL800 DMS(Data Management System)"},{"status":"unaffected","version":"Application software versions >= 6.20MR2 with Windows 10 operating system","versionType":"ABL800 DMS(Data Management System)"}],"platforms":[]},{"vendor":"Radiometer Medical Aps","product":"AQT90 FLEX Analyzers","defaultStatus":"unaffected","versions":[{"status":"affected","version":"Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems","versionType":"AQT90 DMS (Data Management System)"},{"status":"unaffected","version":"Application software versions >= 8.13 MR2 with Windows 10 operating system","versionType":"AQT90 DMS (Data Management System)"}],"platforms":[]}],"packages":[],"cwes":["CWE-287"],"aliases":[],"references":[{"url":"https://www.radiometer.com/myradiometer","tags":[]}],"patchReferenced":false,"fixAvailable":false,"exploitReferenced":false,"kev":false,"kevAdded":null,"kevDue":null,"kevEntry":null,"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"total"},"epss":0.00498,"epssPercentile":0.43616,"sources":[{"name":"CVE List","url":"https://www.cve.org/CVERecord?id=CVE-2025-14097","ok":true},{"name":"NVD","url":"https://nvd.nist.gov/vuln/detail/CVE-2025-14097","ok":false,"note":"unavailable (rate limited?)"},{"name":"OSV","url":"https://osv.dev/vulnerability/CVE-2025-14097","ok":false,"note":"no open-source package data"},{"name":"CISA KEV","url":"https://www.cisa.gov/known-exploited-vulnerabilities-catalog","ok":true,"note":"not listed"},{"name":"EPSS","url":"https://api.first.org/data/v1/epss?cve=CVE-2025-14097","ok":true,"note":"scored 2026-11-20"}],"aiTitle":false,"tech":[],"critical":[],"attention":null}