The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP addr…
Contec Health CMS8000 Patient Monitor
Affected versions
| Product | Affected | Fixed in |
|---|---|---|
| Contec Health CMS8000 Patient Monitor |
| No fix yet |
Original description
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.