Contec Health CMS8000
Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Affected versions
| Product | Affected | Fixed in |
|---|---|---|
| Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor |
| No fix yet |
Original description
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
From the CVE record · icscert