ENDRESS+HAUSER: Ecograph T utilizing Webserver firmware version 1.x has improper privilege management
Endress+Hauser RSG35 - Ecograph T · Endress+Hauser ORSG35 - Ecograph T Neutral/Private Label
Affected versions
| Product | Affected | Fixed in |
|---|---|---|
| Endress+Hauser RSG35 - Ecograph T |
| No fix yet |
| Endress+Hauser ORSG35 - Ecograph T Neutral/Private Label |
| No fix yet |
Original description
Endress+Hauser Ecograph T (Neutral/Private Label) (RSG35, ORSG35) with Firmware version prior to V2.0.0 is prone to improper privilege management. The affected device has a web-based user interface with a role-based access system. Users with different roles have different write and read privileges. The access system is based on dynamic "tokens". The vulnerability is that user sessions are not closed correctly and a user with fewer rights is assigned the higher rights when he logs on.