The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sen…
Baxter Sigma Spectrum Infusion Pumps
Affected versions
| Product | Affected | Fixed in |
|---|---|---|
| Baxter Sigma Spectrum Infusion Pumps |
| No fix yet |
Original description
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.