{"schemaVersion":1,"url":"https://cve.openhack.com/cve/CVE-2020-12040","id":"CVE-2020-12040","title":"Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 3570…","description":"Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the","descriptionLong":"Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.","published":"2020-06-29T13:45:58.000Z","updated":"2026-06-17T02:51:19.680Z","reserved":"2020-04-21T00:00:00.000Z","status":"Modified","rejected":false,"assigner":"icscert","metric":{"version":"3.1","score":9.8,"vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","rating":"critical","source":"NVD"},"metrics":[{"version":"3.1","score":9.8,"vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","rating":"critical","source":"NVD"},{"version":"2.0","score":5,"vector":"AV:N/AC:L/Au:N/C:P/I:N/A:N","rating":"medium","source":"NVD"}],"rating":"critical","products":["Baxter Sigma Spectrum Infusion Pumps"],"affectedProducts":[{"vendor":"n/a","product":"Baxter Sigma Spectrum Infusion Pumps","defaultStatus":null,"versions":[{"status":"affected","version":"Sigma Spectrum v6.x model 35700BAX, Baxter Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module v9,11,13,14,15,16,v20D29,v20D30,v20D31,v22D24, Baxter Spectrum v8.x with Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum LVP v8.x w/Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24"}],"platforms":[]}],"packages":[],"cwes":["CWE-319"],"aliases":[],"references":[{"url":"https://www.us-cert.gov/ics/advisories/icsma-20-170-04","tags":["x_refsource_MISC","x_transferred","Third Party Advisory","US Government Resource"]}],"patchReferenced":false,"fixAvailable":false,"exploitReferenced":false,"kev":false,"kevAdded":null,"kevDue":null,"kevEntry":null,"ssvc":null,"epss":0.00972,"epssPercentile":0.62695,"sources":[{"name":"CVE List","url":"https://www.cve.org/CVERecord?id=CVE-2020-12040","ok":true},{"name":"NVD","url":"https://nvd.nist.gov/vuln/detail/CVE-2020-12040","ok":true,"note":"Modified"},{"name":"OSV","url":"https://osv.dev/vulnerability/CVE-2020-12040","ok":false,"note":"no open-source package data"},{"name":"CISA KEV","url":"https://www.cisa.gov/known-exploited-vulnerabilities-catalog","ok":true,"note":"not listed"},{"name":"EPSS","url":"https://api.first.org/data/v1/epss?cve=CVE-2020-12040","ok":true,"note":"scored 2026-11-19"}],"aiTitle":false,"tech":[],"critical":[],"attention":null}