{"schemaVersion":1,"url":"https://cve.openhack.com/cve/CVE-2019-6538","id":"CVE-2019-6538","title":"Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control","description":"The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia ","descriptionLong":"The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.","published":"2019-03-25T21:26:03.000Z","updated":"2026-06-17T02:39:12.980Z","reserved":"2019-01-22T00:00:00.000Z","status":"Modified","rejected":false,"assigner":"icscert","metric":{"version":"3.1","score":9.3,"vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H","rating":"critical","source":"icscert"},"metrics":[{"version":"3.1","score":9.3,"vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H","rating":"critical","source":"icscert"},{"version":"3.1","score":9.3,"vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H","rating":"critical","source":"hq.dhs.gov"},{"version":"3.1","score":6.5,"vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N","rating":"medium","source":"NVD"},{"version":"2.0","score":3.3,"vector":"AV:A/AC:L/Au:N/C:N/I:P/A:N","rating":"low","source":"NVD"}],"rating":"critical","products":["Medtronic Conexus Radio Frequency Telemetry Protocol","Medtronic MyCareLink Monitor","Medtronic CareLink Monitor","Medtronic CareLink 2090 Programmer","Medtronic Amplia CRT-D","Medtronic Claria CRT-D","Medtronic Compia CRT-D","Medtronic Concerto CRT-D","Medtronic Concerto II CRT-D","Medtronic Consulta CRT-D","Medtronic Evera ICD","Medtronic Maximo II CRT-D","Medtronic Maximo II ICD","Medtronic Mirro ICD","Medtronic Nayamed ND ICD","Medtronic Primo ICD","Medtronic Protecta ICD, Protecta CRT-D","Medtronic Secura ICD","Medtronic Virtuoso ICD","Medtronic Virtuoso II ICD","Medtronic Visia AF ICD","Medtronic Viva CRT-D","Medtronic Brava CRT-D","Medtronic Mirro MRI ICD"],"affectedProducts":[{"vendor":"Medtronic","product":"Conexus Radio Frequency Telemetry Protocol","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"MyCareLink Monitor","defaultStatus":"unaffected","versions":[{"status":"affected","version":"24950"},{"status":"affected","version":"24952"}],"platforms":[]},{"vendor":"Medtronic","product":"CareLink Monitor","defaultStatus":"unaffected","versions":[{"status":"affected","version":"2490C"}],"platforms":[]},{"vendor":"Medtronic","product":"CareLink 2090 Programmer","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Amplia CRT-D","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Claria CRT-D","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Compia CRT-D","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Concerto CRT-D","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Concerto II CRT-D","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Consulta CRT-D","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Evera ICD","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Maximo II CRT-D","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Maximo II ICD","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Mirro ICD","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Nayamed ND ICD","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Primo ICD","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Protecta ICD, Protecta CRT-D","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Secura ICD","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Virtuoso ICD","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Virtuoso II ICD","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Visia AF ICD","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Viva CRT-D","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Brava CRT-D","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]},{"vendor":"Medtronic","product":"Mirro MRI ICD","defaultStatus":"unaffected","versions":[{"status":"affected","version":"All versions"}],"platforms":[]}],"packages":[],"cwes":["CWE-284","CWE-306","CWE-862"],"aliases":[],"references":[{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01","tags":["x_refsource_MISC","x_refsource_CONFIRM","x_transferred","Mitigation","Third Party Advisory","US Government Resource"]},{"url":"http://www.securityfocus.com/bid/107544","tags":["vdb-entry","x_refsource_BID","x_transferred","Third Party Advisory","VDB Entry"]}],"patchReferenced":false,"fixAvailable":false,"exploitReferenced":false,"kev":false,"kevAdded":null,"kevDue":null,"kevEntry":null,"ssvc":null,"epss":0.00895,"epssPercentile":0.60235,"sources":[{"name":"CVE List","url":"https://www.cve.org/CVERecord?id=CVE-2019-6538","ok":true},{"name":"NVD","url":"https://nvd.nist.gov/vuln/detail/CVE-2019-6538","ok":true,"note":"Modified"},{"name":"OSV","url":"https://osv.dev/vulnerability/CVE-2019-6538","ok":false,"note":"no open-source package data"},{"name":"CISA KEV","url":"https://www.cisa.gov/known-exploited-vulnerabilities-catalog","ok":true,"note":"not listed"},{"name":"EPSS","url":"https://api.first.org/data/v1/epss?cve=CVE-2019-6538","ok":true,"note":"scored 2026-11-18"}],"aiTitle":false,"tech":[],"critical":[],"attention":null}