{"schemaVersion":1,"url":"https://cve.openhack.com/cve/CVE-2019-25724","id":"CVE-2019-25724","title":"Dräger Infinity M300 VG2.x Network-Based Denial of Service","description":"Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, te","descriptionLong":"Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.","published":"2026-06-02T19:11:51.652Z","updated":"2026-07-22T19:10:00.120Z","reserved":"2026-06-02T19:05:51.385Z","status":"Awaiting Analysis","rejected":false,"assigner":"VulnCheck","metric":{"version":"4.0","score":7.1,"vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N","rating":"high","source":"VulnCheck"},"metrics":[{"version":"4.0","score":7.1,"vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N","rating":"high","source":"VulnCheck"},{"version":"4.0","score":7.1,"vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","rating":"high","source":"vulncheck.com"},{"version":"3.1","score":6.5,"vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","rating":"medium","source":"VulnCheck"},{"version":"3.1","score":6.5,"vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","rating":"medium","source":"vulncheck.com"}],"rating":"high","products":["Dräger Infinity M300"],"affectedProducts":[{"vendor":"Dräger","product":"Infinity M300","defaultStatus":"unknown","versions":[{"status":"affected","version":"VG2.x","lessThanOrEqual":"VG2.3.1","versionType":"custom"},{"status":"unaffected","version":"VG3.0","versionType":"custom"}],"platforms":[]}],"packages":[],"cwes":["CWE-400"],"aliases":[],"references":[{"url":"https://static.draeger.com/security/download/2019-277-02-M300-VG2x-Security-Advisory.pdf","tags":["vendor-advisory"]},{"url":"https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-x-network-based-denial-of-service","tags":["third-party-advisory"]}],"patchReferenced":false,"fixAvailable":false,"exploitReferenced":false,"kev":false,"kevAdded":null,"kevDue":null,"kevEntry":null,"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"partial"},"epss":0.00175,"epssPercentile":0.07067,"sources":[{"name":"CVE List","url":"https://www.cve.org/CVERecord?id=CVE-2019-25724","ok":true},{"name":"NVD","url":"https://nvd.nist.gov/vuln/detail/CVE-2019-25724","ok":true,"note":"Awaiting Analysis"},{"name":"OSV","url":"https://osv.dev/vulnerability/CVE-2019-25724","ok":false,"note":"no open-source package data"},{"name":"CISA KEV","url":"https://www.cisa.gov/known-exploited-vulnerabilities-catalog","ok":true,"note":"not listed"},{"name":"EPSS","url":"https://api.first.org/data/v1/epss?cve=CVE-2019-25724","ok":true,"note":"scored 2026-11-19"}],"aiTitle":false,"tech":[],"critical":[],"attention":null}