Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface
Dräger Perseus A500
Affected versions
| Product | Affected | Fixed in |
|---|---|---|
| Dräger Perseus A500 |
|
|
Original description
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
From the CVE record · VulnCheck