{"schemaVersion":1,"url":"https://cve.openhack.com/cve/CVE-2019-25719","id":"CVE-2019-25719","title":"Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling","description":"Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can","descriptionLong":"Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.","published":"2026-06-02T14:00:52.578Z","updated":"2026-07-22T19:10:00.120Z","reserved":"2026-06-02T13:54:01.021Z","status":"Awaiting Analysis","rejected":false,"assigner":"VulnCheck","metric":{"version":"4.0","score":8.8,"vector":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N","rating":"high","source":"VulnCheck"},"metrics":[{"version":"4.0","score":8.8,"vector":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N","rating":"high","source":"VulnCheck"},{"version":"4.0","score":8.8,"vector":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","rating":"high","source":"vulncheck.com"},{"version":"3.1","score":8.6,"vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","rating":"high","source":"VulnCheck"},{"version":"3.1","score":8.6,"vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","rating":"high","source":"vulncheck.com"}],"rating":"high","products":["Dräger Infinity Acute Care System","Dräger Standalone Infinity M540 patient monitor"],"affectedProducts":[{"vendor":"Dräger","product":"Infinity Acute Care System","defaultStatus":"unknown","versions":[{"status":"affected","version":"0","lessThanOrEqual":"VG4.1.1","versionType":"custom"},{"status":"affected","version":"0","lessThanOrEqual":"VG4.0.3","versionType":"custom"},{"status":"unaffected","version":"VG4.2","versionType":"custom"}],"platforms":[]},{"vendor":"Dräger","product":"Standalone Infinity M540 patient monitor","defaultStatus":"unknown","versions":[{"status":"affected","version":"0","lessThanOrEqual":"VG4.1.1","versionType":"custom"},{"status":"unaffected","version":"VG4.2","versionType":"custom"}],"platforms":[]}],"packages":[],"cwes":["CWE-924"],"aliases":[],"references":[{"url":"https://static.draeger.com/security/download/PSA-19-255-02_Product-Security-Advisory-IACS-VG4.1.pdf","tags":["vendor-advisory"]},{"url":"https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-via-network-message-handling","tags":["third-party-advisory"]}],"patchReferenced":false,"fixAvailable":false,"exploitReferenced":false,"kev":false,"kevAdded":null,"kevDue":null,"kevEntry":null,"ssvc":{"exploitation":"none","automatable":"yes","technicalImpact":"partial"},"epss":0.00132,"epssPercentile":0.02845,"sources":[{"name":"CVE List","url":"https://www.cve.org/CVERecord?id=CVE-2019-25719","ok":true},{"name":"NVD","url":"https://nvd.nist.gov/vuln/detail/CVE-2019-25719","ok":true,"note":"Awaiting Analysis"},{"name":"OSV","url":"https://osv.dev/vulnerability/CVE-2019-25719","ok":false,"note":"no open-source package data"},{"name":"CISA KEV","url":"https://www.cisa.gov/known-exploited-vulnerabilities-catalog","ok":true,"note":"not listed"},{"name":"EPSS","url":"https://api.first.org/data/v1/epss?cve=CVE-2019-25719","ok":true,"note":"scored 2026-11-19"}],"aiTitle":false,"tech":[],"critical":[],"attention":null}