Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explore…
ICS-CERT Dräger Infinity Delta
Affected versions
| Product | Affected | Fixed in |
|---|---|---|
| ICS-CERT Dräger Infinity Delta |
| No fix yet |
Original description
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.
From the CVE record · icscert