An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Vers…
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump
High8.1
Affected versions
| Product | Affected | Fixed in |
|---|---|---|
| Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump |
| No fix yet |
Original description
An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump does not require authentication if the pump is configured to allow FTP connections.