Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq I…
Hospira Plum A+ Infusion System · Hospira Plum A+3 Infusion System · Hospira Symbiq Infusion System
Affected versions
| Product | Affected | Fixed in |
|---|---|---|
| Hospira Plum A+ Infusion System |
| No fix yet |
| Hospira Plum A+3 Infusion System |
| No fix yet |
| Hospira Symbiq Infusion System |
| No fix yet |
Original description
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.